(2017) Fabrication of an amperometric glucose biosensor based on a Prussian blue/carbon nanotube/ionic liquid modified glassy carbon electrode. Journal of Applied Biotechnology Reports. pp. 603-608. ISSN 23221186 (ISSN)
Text
Fabrication of an amperometric glucose biosensor based on a Prussian bluecarbon nanotubeionic liquid modified glassy carbon electrode.pdf Download (630kB) |
Abstract
An amperometric glucose biosensor was developed based on synergistic contributions of prussian blue (PB) and a bucky gel (BG) consisting of carbon nanotubes (CNTs) and ionic liquid (IL). The PB nanoparticles were first deposited onto the surface of a BG modified glassy carbon (GC) electrode (BG/GC). Then, the Ni2+ ions were electrochemically inserted into the PB lattice to improve its stability at physiological pH. Afterwards, Glucose oxidase (GOx) was immobilized on the BG/GC electrode using a cross-linking method. Amperometric measurements of glucose were performed at -0.05 V vs. Ag/AgCl in 0.05 M phosphate buffer solution at pH 7.4. The glucose biosensor exhibited a sensitivity of 45.03 µA/(mM.cm2) with a detection limit of 5x10-7 M. The amperometric response was linear in the range of 5x10-7 to 8.3x10-4M. © 2017, Baqiyatallah University of Medical Sciences. All rights reserved.
Item Type: | Article |
---|---|
Keywords: | Biosensore Carbon nanotube Glucose Ionic liquid Nanoparticle |
Divisions: | |
Page Range: | pp. 603-608 |
Journal or Publication Title: | Journal of Applied Biotechnology Reports |
Journal Index: | Scopus |
Volume: | 4 |
Number: | 2 |
ISSN: | 23221186 (ISSN) |